3.5.1 Change of active measurement channel
• in order to change currently active measurement channel press
the
keys in channel 1 display field.
3.5.2 Measurement channel tuning procedure for the PT-100 input.
ATTENTION: Check position of measurement input type switch on the back panel
of the Recorder.
1 - select channel to tune by pressing the keys in channel 1
display field.
2 - preset model signal value (0°C; Ro=100 Ω) or place sensing element in water
and ice mix, which permits additional correction of sensing element error.
3 - carry out correction of zero by pressing the key in channel 1 display
field.
(in order to cancel the correction press the
key in channel 2 display field)
completed correction will be indicated by blinking of the diode built-in in the
key in channel 2 display field.
4
- preset model signal value in a selected point of the PT-100 sensor characteristic
curve.
Recommended values are:
- R100 = 138.5 Ω - which corresponds to the indication of 100 °C
then, by turning potentiometers (for respective channels
Ch1,Ch2,Ch3,Ch4,Ch5,Ch6,Ch7,Ch8) on the back panel of the Recorder, ensure
that correct read out value is being displayed.